FECO CASHTILL # 5

FECO: CASHTILL # 5
MODIFICATION No: CASH NO 1
PRODUCT: RS232 CASHDRAWER
SUB-ASSEMBLY N/A
ESTIMATED MAN HOURS ONE
CLASS OF CHANGE: MANDATORY BY FIELD SERVICE AT NEXT VISIT

MANDATORY BY PRODUCTION ON COMMISSIONING

MANDATORY BY REPAIRS ON FAILURE OR REWORK.

REASON FOR CHANGE: TO OVERCOME THE PROBLEM OF SOLENOID BURN OUT.

1.Reason

        Problems have arisen with RS232 cash drawers where the energising signal has remained high for long
        enough to cause the solenoid to burn out.

2.Change

        To overcome this problem, the cash drawer is to be rewired so that the micro switch is in series with the
        solenoid. See fig 1 & 2.

3.new operation

3.1.energising signal at pin 2 goes high.

3.2.the diode bridge gives 24 volt output to the solenoid, via the closed contacts of the micro switch. The
        solenoid energises.

3.3.the cash drawer opens, opening the micro switch, removing the power from the solenoid, so protecting it.

3.4.if the cash drawer is closed, whilst the energising signal at pin 2 is still high, the solenoid will re-energise and
        open it again, then de-energise.

        (tech serv ref: 23/2/3)

4.circuit diagrams

5.figure 1 shows the existing circuit diagram, figure 2 shows the modified diagram.

        The circuits are shown with the cash drawer closed and the energising signal at pin 2 being low, i.e. not
        present.

6.to modify

6.1.ensure the cash drawer is disconnected from the mains and data supply.

6.2.open the cash drawer, remove the cash tray and the frame with the front panel on. Also remove the top
        panel of the cash drawer, by removing the four nuts from the studs inside the drawer. This gives access.

6.3.inside the cashdrawer at the rear, is the black control unit on the left and the solenoid mechanism on the
        right. See figure 3.

        The left side of the solenoid plunger is connected to the frame latching mechanism, with the microswitch
        below it.

6.4.the four way connector block is mounted above the solenoid and is called 1 to 4, left to right, for indentity
        purposes. The top connector is called t and the bottom b. See figure 4.

        Note: the cable colours may not be identical.

6.5.the existing wiring to the connector block should be:

        Microswitch connected to 1b & 2b

        (wires black & blue)

        Solenoid connected to 3b & 4b

        (wires both red

        Solenoid drive connected to 3t & 4t

        (wires red & blue)

        Rs232 connector pins 7 & 20 connected to 1t & 2t

        (wires black & white)

6.6.disconnect lead from 1t and connect it to 2t. There should now be two leads connected to 2t.

6.7.disconnect lead from 4t and connect it to 1t.

6.8.disconnect lead from 2b and connect it to 4t.

6.9.see figure 5.

6.10replace the frame, but not the cash tray or top of the cash drawer; and test the unit.

7.testing

7.1.ensure the cashdrawer is in the open position, observe the solenoid in it's de-energised state.

7.2.connect the cash drawer to the data and mains supply.

7.3.observe the solenoid. Ensure that it is still de-energised.

7.4.with the cash drawer in the open position, enter:

        Protocol n c=0 where n is the process number

        And ensure that the solenoid is still de-energised.

7.5.now close the cash drawer and ensure that the frame latching mechanism holds the drawer shut.

7.6.enter protocol n c=1 and ensure cash drawer opens and solenoid de-energises.

7.7.try closing the cashdrawer, it should vibrate and open immediately.

7.8.enter protocol n c=0

7.9.close the cash drawer and ensure that it stays shut.

7.10.replace the cash drawer top panel and cash tray.

8.recording action

        Affix a modification label adjacent to the serial number plate. Strike through but do not obliterate 1.

Return to Index